Kryo 1060 - 180

Dimensions

Internal

External

Height

64 cm

112 cm

Width

1 x 50 cm

86 cm

Depth

50 cm

116 cm

Weight

100 kg (shipping weight inc. packaging) approx.

Capacity

3500 x 2ml ampoules

 

Kryo 1060 - 380

Dimensions

Internal

External

Height

64 cm

 112 cm

Width

2 x 50 cm

132 cm

Depth

50 cm

116 cm

Weight

160 kg (shipping weight inc. packaging) approx.

Capacity

8000 x 2ml ampoules

 

Circulation

Horizontal laminar flow

Temperature range

+40șC to -100șC. Warning! The freezer is fitted with a manually-resettable thermal cut-out to prevent over heating. This will operate if the chamber is programmed to run above +40șC and will require a Service Engineer to reset it.

Cooling medium

Liquid nitrogen 22 - 30 psi

Heater

1700W

Sensors: Control and sample

4-wire Platinum resistance thermometer. Sensors are linearised in software to international standards that utilise a 4096-point lookup table based on BS1904:1984 Table 1. Calibration facility provided.

Accuracy

±0.5șC at a hold at 0șC (dynamic accuracy depends on actual program, e.g. rate of change of temperature

Heating rates

0.01șC/min to 1șC/min

Cooling rates

-0.01șC/min to -5șC/min

Power requirements

230 V ~ 50/60Hz, 16A (max.) The freezer may be damaged by voltage surges in excess of 15% above nominal

Chart recorder

Sensitivity: 16.7mV/șC.
Nominal impedance > 10K

Standards

Designed to comply with BSEN 61010,  EN50082-2, EN50081-2

Storage temperature

-10șC to +70șC

Storage humidity

Up to 95% non-condensing

Operating temperature

5șC to 40șC

Operating humidity

90%