Measurement |
|||
|
Range |
-200°C to 0°C |
||
|
PC Programme |
Display units user selectable -°C, °F or K |
||
|
Probe Type |
Platinum Resistance, Pt 100, Class B |
||
|
Resolution |
Approximately 1.0°C over whole operating range |
||
|
Accuracy |
±1.5°C at 0°C, rising to ±2.5°C at -200°C, including uncer-tainty due to measurement resolution and probe errors |
||
|
Calibration |
Traceable to NAMAS standards; semi-automatic user calibration possible using optical PC calibration pack; calibration certificate printed automatically |
||
|
Data Logging |
|||
|
Number of points |
8192 |
||
|
Logging Interval |
30 seconds to 99 minutes in 30 second increments |
||
|
Logging Resolution |
0.1°C |
||
|
Max Number of Logs |
8192 (Limited to number of points) |
||
|
Alarm Generation |
|||
|
High Alarm Level |
-200°C to 0°C in 1.0°C steps |
||
|
Low Alarm Level |
-200°C to 0°C in 1.0°C steps |
||
|
Alarm Delay Period |
0 to 120 logging intervals |
||
|
Control & Indication |
|||
|
Manual Control |
Magnetic non contact Start / Stop Switch using supplied fob |
||
|
Communication |
RS232 Serial communications to proprietary PC program |
||
|
Programming |
Using propriety PC program |
||
|
Indicators |
Three x LED Indicators: Yellow=Normal, Orange=Warning, Red=Alarm |
||
|
Battery |
|||
|
Type |
Lithium Thionyl Chloride, internally fitted, factory replaceable only |
||
|
Life |
Approximately 3 years in normal operation; battery life remain-ing is indicated via PC programme |
||
|
PC Software Requirements |
PC to be Windows ‘95, ‘98 or NT |
||
|
Ordering Information |
|||
|
Description |
Ordering Ref |
||
|
ShipsLog with temperature probe |
GDSHIPSLOG-FL |
||
|
ShipsLog with fixed probe 720mm long |
GDSHIPSLOG-FL-720 |
||
|
ShipsLog for Cryoshipper vapour shipper |
GDSHIPSLOG-CS |
||
|
ShipsLog for SC4/3V vapour shipper |
GDSHIPSLOG-SC4/3V |
||
|
ShipsLog for SC4/2V vapour shipper |
GDSHIPSLOG-SC4/2V |
||
|
Temperature Probe for ShipsLog |
GDSHIPSLOG-PRT-610MM |
||
|
Serial Cable to connect ShipsLog to PC |
AC013808 |
||
|
Please refer to your vessel documentation for details. Note that use of the ShipsLog for the SC4/3V could re-sult in an increase in liquid nitrogen evaporation rate of 22% and this may affect the hold-up time of the system. |
|||
|
|
|
||
Planer Controlled Rate Freezer Kryo 360 1.7 560 16 750 30 Integra Cryogenic Low Temperature Freezing Cryopreservation storage vessel laboratory Controlled Rate Freezers Controlled rate freezer controlled rate freezer controlled rate freezer